Reason about SaMD (Software as a Medical Device) regulatory compliance for FDA, EU MDR, and global submissions. Use when the user asks about IEC 62304 safety classification, ISO 14971 risk management, 21 CFR 820/QMSR quality system requirements, Design History File structure, 510(k)/De Novo/PMA pathway selection, SOUP/OTS assessment, verification and validation planning, cybersecurity for medical devices, PCCP for AI/ML devices, design reviews, traceability matrices, or post-market surveillance. Triggers include "SaMD", "medical device software", "IEC 62304", "ISO 14971", "ISO 13485", "21 CFR 820", "QMSR", "510(k)", "De Novo", "PMA", "design history file", "DHF", "software safety classification", "SOUP list", "risk management", "hazard analysis", "design review", "V&V protocol", "FDA submission", "EU MDR", "clinical evaluation", "PCCP", "predetermined change control", "GMLP", "cybersecurity premarket", "Part 11", "design controls", "design inputs", "design outputs", "traceability matrix", "post-market surveillance", "CAPA", "Class II device", "Class III device", "regulatory pathway", "predicate device", "substantial equivalence".

SKILL.md

SaMD Compliance — Reasoning Skill

Overview

You are an expert in Software as a Medical Device (SaMD) regulatory compliance. When the user asks about FDA software device submissions, IEC 62304 lifecycle, ISO 14971 risk management, design controls, or global regulatory strategy, apply the decision frameworks below.

Usage

  • Invoke when evaluating SaMD regulatory compliance for FDA, EU MDR, or multi-market submissions
  • Use when planning design controls, risk management, V&V, or submission strategy
  • Activate for safety classification, SOUP assessment, or AI/ML device lifecycle questions

Response Format

  • Lead with the direct recommendation or classification (≤3 sentences)
  • Structure as: recommendation → justification (citing specific standards/clauses) → caveats
  • Use tables for comparisons; bullet points for criteria lists
  • Omit background the user already knows — they asked the question
  • Target: 200-400 words unless the user requests exhaustive detail

The decision trees and frameworks in this skill are for internal reasoning only. Apply them to reach your conclusion, but do not reproduce them in your response. Present only the final recommendation with supporting evidence.


1. IEC 62304 Safety Classification Decision Tree

Does the software system contribute to a hazardous situation?
├─ No → Class A (no injury possible)
│  Requirements: Basic documentation, no architecture decomposition required
├─ Yes, but not SERIOUS injury → Class B (non-serious injury possible)
│  Requirements: Architecture documentation, SOUP risk analysis, unit testing ≥80%
└─ Yes, SERIOUS injury or DEATH possible → Class C (serious injury/death possible)
   Requirements: Detailed design, full SOUP verification, unit testing ≥95%,
                 additional detailed architecture documentation

Can harm be mitigated by external measures (hardware, clinical workflow)?
├─ Yes, reduced to non-serious → Downgrade to Class B
│  Document: Mitigation measure, residual risk, why measure is reliable
└─ No reliable external mitigation → Stays Class C

Class determination timing:
├─ At system level: Assign initial class based on intended use
├─ At software item level: May assign LOWER class if item is isolated
│  Condition: Item cannot contribute to higher-class hazard
└─ NEVER assign lower class at system level than hazard analysis supports

2. FDA Regulatory Pathway Decision Tree

Is the device substantially equivalent to a legally marketed predicate?
├─ Yes, same intended use + same/different technology (equiv. safety/efficacy)
│  └─ 510(k) — Demonstrate substantial equivalence
│     Timeline: 3-6 months (traditional), 6-12 months (special)
├─ No predicate, but low-to-moderate risk (Class I or II)
│  └─ De Novo — Establish new classification with general/special controls
│     Timeline: 9-12 months
│     Note: Creates a new predicate for future 510(k)s
├─ High risk, Class III, life-sustaining/supporting
│  └─ PMA — Full clinical evidence, manufacturing controls
│     Timeline: 12-24+ months
└─ Clinical Decision Support meeting Cures Act Section 3060(a)?
   All 4 criteria met:
   1. Not intended to acquire/process/analyze medical images/signals
   2. Intended for healthcare professionals (not patients)
   3. Intended to enable HCP to independently review basis
   4. Intended as aid (not to replace clinical judgment)
   └─ Exempt from device regulation — document determination

3. Design Controls Lifecycle (21 CFR 820.30 / QMSR)

PhaseKey OutputIEC 62304 MappingBlocking Gate
User NeedsUser Needs Document§5.2 Software requirements processNone
Design InputSRS (Software Requirements Spec)§5.2.1-5.2.6 RequirementsDR1
Design OutputArchitecture + Detailed Design§5.3 Software architectural designDR2
VerificationTest protocols + reports§5.5 Software integration testing, §5.6 System testingDR3
ValidationClinical validation evidence§5.7 Software releaseSubmission
TransferManufacturing/deployment procedures§5.8 Maintenance processRelease
ChangesDesign change assessment§5.2 (re-entry)Per-change

Critical rule: Design Reviews (DR1, DR2, DR3) are BLOCKING for Class B/C devices. Each requires documented attendees, findings, and formal sign-off.

4. SOUP/OTS Assessment Framework

Risk-Based SOUP Classification

Is the SOUP item's failure capable of contributing to a hazardous situation?
├─ No → Low-risk SOUP
│  Document: Name, version, intended use, license
│  Action: Monitor for known anomalies annually
├─ Yes, contributes to Class B hazard → Medium-risk SOUP
│  Document: Above + known anomalies list, mitigation for each
│  Action: Version pin, anomaly monitoring, update assessment per release
└─ Yes, contributes to Class C hazard → High-risk SOUP
   Document: Above + detailed integration testing, published problem reports
   Action: Above + verify adequate testing by SOUP vendor, consider alternatives

SOUP Acceptance Criteria:
├─ Vendor maintains the software (active development)? → Preferred
├─ Known anomaly list published? → Required for Class B/C
├─ CVE history acceptable? → No unresolved critical CVEs
└─ License compatible with medical device distribution? → Required

SOUP vs Custom Decision

FactorUse SOUPBuild Custom
Well-validated open source library
Critical safety function with no validated library
Standard utility (logging, HTTP, encryption)
Regulatory precedent for the library
SOUP vendor unresponsive to anomaly reports
Time-to-market critical, library mature

5. ISO 14971 Risk Management Process

Severity Classification

LevelDescriptionExamplesAcceptable Probability
S1NegligibleInconvenience, no injuryAny
S2MinorTemporary minor injuryOccasional
S3SeriousInjury requiring interventionRemote
S4CriticalPermanent impairmentImprobable
S5CatastrophicDeathImprobable

Risk Control Priority (ISO 14971 §7.1)

Risk exceeds acceptability threshold?
├─ Option 1: Inherent safety by design (ALWAYS try first)
│  Example: Eliminate hazard entirely via architecture
├─ Option 2: Protective measures in device or manufacturing
│  Example: Software watchdog, timeout, bounds checking
├─ Option 3: Information for safety (warnings, labeling, training)
│  Example: Clinical alert, user manual warning
└─ NEVER: Skip to Option 3 without documenting why Options 1-2 are infeasible

Residual Risk Assessment:
├─ Individual residual risk acceptable? → Document and proceed
├─ Individual unacceptable but reducible? → Apply additional controls
└─ Overall residual risk vs. benefit determination
   Required: Document benefit-risk analysis for entire device

Hazard Analysis Common Pitfalls

  1. Wrong: Listing software bugs as hazards Right: Trace bug → hazardous situation → harm. The HARM is what matters.
  2. Wrong: Copying generic hazard lists without device-specific analysis Right: Analyze YOUR device's specific failure modes in YOUR clinical context
  3. Wrong: Assuming software cannot cause physical harm Right: Delayed diagnosis, wrong treatment recommendation, alert fatigue ALL cause harm
  4. Wrong: Risk controls that rely entirely on user vigilance Right: Design-level controls first; training/warnings are last resort

6. V&V Planning by Safety Class

ActivityClass AClass BClass C
Requirements traceability
Unit testingOptional≥80% coverage≥95% coverage
Integration testingOptional✅ + fault injection
System testing✅ + stress/boundary
Regression testingOn changeFull suiteFull suite + risk-based
Clinical validationIf claimedIf claimed✅ Mandatory
Cybersecurity testingIf connected✅ Penetration test✅ Full threat model
Usability validationIf HCP user✅ Formative✅ Summative

V&V Failure Recovery Decision Tree

Test fails acceptance criteria
├─ Safety-related requirement?
│  ├─ Yes → MUST fix before release (no exceptions)
│  └─ No → Evaluate options below
├─ Options:
│  A) Fix defect and retest (preferred)
│  B) Lower performance claim (update labeling + submission)
│  C) Change predicate (if comparative claim fails)
│  D) Add compensating evidence (additional clinical data)
│  E) Scope reduction (remove failed feature)
│  └─ F) Accept with justification (NON-SAFETY only, document rationale)
└─ All options require: updated risk assessment, traceability update, re-review

7. Cybersecurity Requirements (FDA 2023 Guidance)

Threat Modeling Minimum Scope

CategoryMust AddressExample Threats
ConfidentialityPHI exposureUnencrypted API, debug endpoints
IntegrityData/algorithm tamperingModel poisoning, input manipulation
AvailabilityDenial of serviceResource exhaustion, dependency failure
AuthenticationUnauthorized accessDefault credentials, session hijacking
Update mechanismSecure patchingUnsigned updates, rollback attacks

SBOM Requirements (per FDA Refuse-to-Accept checklist)

  • Format: CycloneDX 1.5+ or SPDX 2.3+
  • Must include: ALL direct + transitive dependencies
  • Per component: name, version, supplier, license, known CVEs
  • Update frequency: Every release + within 24h of critical CVE disclosure
  • FDA will RTA (Refuse to Accept) submissions without SBOM as of Oct 2023

8. AI/ML SaMD Lifecycle (PCCP Framework)

FDA Total Product Lifecycle for AI/ML

Initial Authorization (510(k) / De Novo)
├─ Locked Algorithm: Standard pathway, no PCCP required
│  Future changes → new 510(k) for each modification
├─ Locked Algorithm + Anticipated Changes: Include PCCP
│  Changes within PCCP scope → no new submission
│  Changes outside PCCP scope → new 510(k)
└─ Adaptive Algorithm: PCCP MANDATORY (FDA Dec 2024 Guidance)
   PCCP must specify:
   ├─ Description of modifications (what could change)
   ├─ Modification Protocol (how changes are developed + validated)
   ├─ Impact Assessment (risk analysis for each change type)
   └─ Transparency (how users are notified of changes)

Good Machine Learning Practice (GMLP) — 10 Principles

#PrincipleVerification Method
1Multi-disciplinary teamDocument team roles (clinical + engineering + regulatory)
2Good Software Engineering PracticeIEC 62304 compliance
3Representative clinical dataDataset demographics vs. intended population
4Independent training/test/validation setsNo data leakage verification
5Reference datasets (ground truth)Adjudication process documented
6Model design fits intended useArchitecture justification document
7Clinically relevant performance metricsSensitivity/Specificity/PPV/NPV/AUC
8Testing across patient subgroupsBias analysis across demographics
9Clear user informationLabeling re: capabilities + limitations
10Deployed model monitoringDrift detection + performance tracking

Performance Monitoring Thresholds

MetricAction ThresholdEscalation
Overall performance (AUC/F1)>5% degradation from validationInvestigate root cause
Subgroup performance>10% gap vs. majority subgroupBias review
False negative rate (safety)Any increase >2%Immediate CAPA
Data distribution shiftOOD rate >15% of inputsRetraining evaluation
Alert volume (CDS devices)>20 alerts/clinician/shiftAlert fatigue review

9. Traceability Requirements

Bidirectional Traceability Matrix Structure

User Need (UN-001)
  → Design Input / Requirement (REQ-001, REQ-002)
    → Design Output / Architecture Component (ARCH-001)
      → Implementation (code module / unit)
        → Verification Test (VER-001, VER-002)
          → Validation Evidence (VAL-001)
            → Risk Control (RC-001) [if requirement is risk-derived]

Traceability Completeness Rules

CheckBlocking?Description
Orphan requirement (no parent UN)YesEvery REQ must trace to a user need
Unverified requirementYesEvery REQ must have ≥1 verification test
Unimplemented risk controlYesEvery RC in hazard analysis must trace to code
Dead code (no tracing requirement)WarningMay indicate scope creep or incomplete specs
Missing validation linkClass B/CEach user need must trace through to validation

10. Design Review Checklist (DR1 / DR2 / DR3)

DR1 — Inception Exit (Before Construction)

ItemCriteria
User needs documentedAll clinical needs captured with acceptance criteria
Requirements completeSRS covers all 9 categories (functional, performance, interface, cybersecurity, usability, risk control, data integrity, PHI, regulatory)
Risk management planISO 14971 plan approved, initial hazard analysis complete
Regulatory strategyPathway selected, predicates identified
Traceability (initial)UN → REQ links established

DR2 — Pre-Validation (After Construction, Before V&V)

ItemCriteria
Architecture documentedIEC 62304 §5.3 satisfied
SOUP assessedAll 3rd-party items risk-classified
Code reviews completeAll units reviewed, findings resolved
Design freezeNo more code changes during V&V
Risk controls implementedAll controls in hazard analysis coded + unit tested
SBOM generatedCycloneDX/SPDX with all dependencies

DR3 — Pre-Submission (After V&V)

ItemCriteria
All tests pass0 open failures on safety/efficacy tests
Coverage metClass-appropriate thresholds achieved
Residual risk acceptableBenefit-risk documented and favorable
Traceability closedFull bidirectional, no orphans
Anomalies dispositionedAll defects resolved or risk-accepted
Labeling reviewedIntended use, contraindications, IFU complete

11. FDA vs EU MDR Comparison for SaMD

AspectFDA (US)EU MDR 2017/745
ClassificationRisk-based (Class I/II/III)Rule 11: SaMD is IIa minimum; IIb/III if serious
Pathway510(k), De Novo, PMANotified Body conformity assessment
Clinical evidenceSubstantial equivalence (510k) or clinical studies (PMA)Clinical Evaluation Report ALWAYS required
Post-marketAnnual reports, MDRsPMCF + PSUR annually + vigilance
AI/ML updatesPCCP frameworkNotified Body re-assessment for significant changes
CybersecurityMandatory SBOM + threat model (2023)MDCG 2019-16 guidance (recommended)
QMS standard21 CFR 820 / QMSR (aligned with ISO 13485)ISO 13485 certification required
Software lifecycleIEC 62304 (recognized consensus standard)IEC 62304 (harmonized standard)
Timeline510(k): 3-6mo; De Novo: 9-12mo; PMA: 12-24moNB audit cycle: 12-18 months
Labeling21 CFR 801 + unique labeling requirementsAnnex I GSPR + SSCP (for implants/Class III)

Key gotcha: EU MDR Rule 11 classifies most SaMD as IIa minimum (vs FDA Class II). SaMD providing information for diagnosis of serious conditions = Class IIb or III in EU even if Class II in US. Plan for the higher classification upfront if dual-market.

12. Common SaMD Compliance Mistakes (Severity-Ranked)

  1. Wrong: Treating IEC 62304 safety class as equivalent to FDA device class Right: IEC 62304 Class A/B/C is SOFTWARE safety; FDA Class I/II/III is DEVICE risk. A Class II device can have Class C software. Why: Incorrect classification leads to insufficient documentation and testing — Critical severity
  2. Wrong: Starting code before design inputs are documented Right: Document user needs → derive requirements → get DR1 approval → THEN build Why: 21 CFR 820.30(c) requires design input documentation before design output. FDA will cite this. — Critical severity
  3. Wrong: Omitting SOUP from the risk analysis Right: Every SOUP item must be assessed for known anomalies that could contribute to hazards Why: IEC 62304 §7.1.3 mandates SOUP risk assessment; unassessed SOUP = undocumented risk — High severity
  4. Wrong: Using "testing" as the only risk control Right: Testing VERIFIES a risk control works — it is not itself a control. Controls are architectural (watchdogs, bounds checks, redundancy). Why: Confusing verification with mitigation leaves residual risk unaddressed — High severity
  5. Wrong: No traceability between risk controls and verification tests Right: Each risk control in the hazard analysis must link to a specific test proving it works Why: Without this link, you cannot demonstrate risk controls are effective — submission deficiency — High severity
  6. Wrong: Performing design reviews as rubber-stamp exercises Right: Design reviews must have documented attendees, specific findings, action items, and resolution evidence Why: FDA inspectors look for "objective evidence" of review — generic sign-offs get 483s — High severity
  7. Wrong: Assuming De Novo is slower/harder than 510(k) for novel devices Right: De Novo creates a new predicate, establishing your market category. Average review: 9-12 months. Why: Forcing substantial equivalence to a poor predicate leads to Additional Information requests and delays — Medium severity
  8. Wrong: Treating cybersecurity as optional for non-connected devices Right: FDA's 2023 guidance applies to ALL devices with software. Threat model scope includes supply chain, update mechanism, and data-at-rest. Why: FDA will RTA submissions without cybersecurity documentation — Medium severity

13. Part 11 (Electronic Records) Decision Tree

Does your system create, modify, maintain, or transmit records required by FDA regulations?
├─ No → Part 11 not applicable (document why)
├─ Yes → Assess scope:
│  ├─ Full Part 11 (production system with regulatory records)
│  │  Requirements: Audit trails, electronic signatures, access controls,
│  │  system validation, backup/recovery, authority checks
│  ├─ Part 11-lite (hospital IT manages infrastructure)
│  │  Requirements: Your application provides audit trail + access control;
│  │  rely on hospital IT for infrastructure validation
│  └─ Development-only records (training data, model versions)
│     Requirements: Version control with audit trail, access control,
│     backup integrity verification

Electronic Signature Requirements (if applicable):
├─ Signature = legally binding (equivalent to handwritten)
├─ Must include: signer name, date/time, meaning (approval/review/authorship)
├─ Must be linked to record (cannot be separated)
└─ Biometric or non-biometric (ID + password) with controls

14. Post-Market Surveillance Essentials

Mandatory Activities by Market

ActivityFDAEU MDRFrequency
Medical Device Report (MDR/vigilance)Within 30 days (FDA) / immediately for serious (EU)
Annual reportAnnually
Periodic Safety Update Report (PSUR)Annually (Class IIa+)
Post-Market Clinical Follow-up (PMCF)Ongoing
Complaint handlingPer event
CAPA processPer finding
Trend analysisQuarterly minimum
Software update reportingPer update (risk-based)

CAPA Trigger Decision Tree

Signal detected (complaint, trend, field event)
├─ Safety-related?
│  ├─ Yes → Immediate CAPA + potential field corrective action
│  └─ No → Evaluate below
├─ Systematic (affects multiple units/users)?
│  ├─ Yes → CAPA (root cause likely systemic)
│  └─ No → Correction only (isolated incident)
├─ Recurring (≥3 occurrences same root cause)?
│  └─ Yes → CAPA mandatory (correction alone is insufficient)
└─ Regulatory requirement (audit finding, FDA warning letter)?
   └─ Yes → CAPA mandatory with defined timeline

When NOT to Use This Skill

  • General software engineering without medical device claims — design controls add overhead without regulatory value
  • Clinical data standards (CDISC SDTM/ADaM) — use cdisc-compliance skill instead
  • Wellness/fitness apps with NO diagnostic or therapeutic claims — not regulated as devices
  • eCTD submission assembly or eSTAR portal mechanics — this skill covers device compliance, not submission logistics
  • Manufacturing quality (GMP, process validation for physical devices) — this skill focuses on software lifecycle

When to Escalate to Human Expert

  • Safety classification disagreement where reasonable arguments support multiple classes — requires cross-functional risk team decision
  • Predicate selection for 510(k) where no clear substantial equivalence exists — may need pre-submission meeting with FDA
  • Clinical evidence sufficiency questions — requires clinical affairs and biostatistics expertise
  • EU MDR classification under Rule 11 for borderline CDS/wellness software — requires Notified Body pre-assessment
  • Post-market safety signal assessment for potential field action — requires medical director and regulatory affairs